Product characteristics |
Number of inputs and outputs |
Number of analogue outputs: 2 |
Measuring range |
|
Process connection |
threaded connection G 1 1/4 flat seal |
Application |
Special feature |
Gold-plated contacts |
Application |
for industrial applications |
Installation |
connection to pipe by means of an adapter |
Media |
water; glycol solutions; coolants; oil |
Note on media |
low-viscosity oils with viscosity: 7…40 mm²/s (40 °C) |
high-viscosity oils with viscosity: 30…68 mm²/s (40 °C) |
|
Medium temperature [°C] |
-10…80 |
Pressure rating [bar] |
16 |
Pressure rating [Mpa] |
1.6 |
Electrical data |
Operating voltage [V] |
19…30 DC; (according to EN 50178 SELV/PELV) |
Current consumption [mA] |
100 |
Min. insulation resistance [MΩ] |
100; (500 V DC) |
Protection class |
III |
Reverse polarity protection |
yes |
Power-on delay time [s] |
10 |
Inputs / outputs |
Number of inputs and outputs |
Number of analogue outputs: 2 |
Outputs |
Total number of outputs |
2 |
Output signal |
analogue signal |
Number of analogue outputs |
2 |
Analogue current output [mA] |
4…20; (scalable) |
Max. load [Ω] |
500 |
Short-circuit protection |
yes |
Type of short-circuit protection |
pulsed |
Overload protection |
yes |
Measuring/setting range |
Measuring range |
|
Display range |
|
Resolution |
|
Analogue start point ASP |
|
Analogue end point AEP |
40…200 l/min |
10.28…52.84 gpm |
|
Max. flow rate [l/min] |
220 |
In steps of |
|
Temperature monitoring |
Measuring range [°C] |
-10…80 |
Resolution [°C] |
0.2 |
Analogue start point [°C] |
-10…62 |
Analogue end point [°C] |
8…80 |
In steps of [°C] |
0.2 |
Accuracy / deviations |
Flow monitoring |
Accuracy (in the measuring range) |
< ± (3 % MW + 0,2 % MEW) / < ± (8 % MW + 0,5 % MEW); (water; glycol: 35%; oil: viscosity 68 mm²/s at 40 °C) |
Repeatability |
1 l/min; 60 l/h; 0,05 gpm; 3 gph |
Temperature monitoring |
Accuracy [K] |
± 3 (Q > 20 l/min) |
Response times |
Flow monitoring |
Response time [s] |
0.25; (dAP = 0) |
Damping for the switching output dAP [s] |
0…1 |
Temperature monitoring |
Dynamic response T05 / T09 [s] |
T09 = 30 (Q > 20 l/min); (water) |
Operating conditions |
Ambient temperature [°C] |
-10…60 |
Storage temperature [°C] |
-25…80 |
Protection |
IP 67 |
Mechanical data |
Weight [g] |
1883 |
Materials |
housing: AlMgSi0,5 anodised; Sealing: FKM; connector housing: brass Optalloy-plated; PA 6.6; cover film: PA |
Materials (wetted parts) |
stainless steel (1.4404 / 316L); FKM; PPS; Centellen 200 |
Process connection |
threaded connection G 1 1/4 flat seal |
Displays / operating elements |
Display |
Display unit |
6 x LED, green (l/min, m³/h, gpm, gph, °C, °F) |
measured values |
alphanumeric display, 4-digit |
programming |
alphanumeric display, 4-digit |
|
Display unit |
l/min; m³/h; gpm; gph; °C; °F |
Accessories |
Accessories (supplied) |
|
Accessories (optional) |
adapter for pipe: 1 x R 1, stainless steel, E40205 |
adapter for pipe: 1 x 1″ NPT, stainless steel, E40206 |
|
Remarks |
Remarks |
sealing: only with supplied Centellen seals |
MW = measured value |
MEW = Final value of the measuring range |
|
Pack quantity |
1 pcs. |
Electrical connection |
Connection |
Connector: 1 x M12; Moulded body: brass, Optalloy-plated; Contacts: gold-plated |