Product characteristics |
Number of inputs and outputs |
Number of digital outputs: 2; Number of analogue outputs: 1 |
Measuring range |
0.5…143.9 m/s |
8…2110 m³/h |
0.12…35.18 m³/min |
|
Process connection |
threaded connection G 1 internal thread |
Application |
Installation |
Adjustable to inside pipe diameters; (38…254 mm) |
Media |
compressed air |
Medium temperature [°C] |
0…60 |
Pressure rating [bar] |
16 |
Pressure rating [Mpa] |
1.6 |
Pressure rating [psi] |
232 |
MAWP (for applications according to CRN) [bar] |
16 |
Electrical data |
Operating voltage [V] |
18…30 DC; (according to EN 50178 SELV/PELV) |
Current consumption [mA] |
< 110 |
Protection class |
III |
Reverse polarity protection |
yes |
Power-on delay time [s] |
1 |
Inputs / outputs |
Number of inputs and outputs |
Number of digital outputs: 2; Number of analogue outputs: 1 |
Outputs |
Total number of outputs |
2 |
Output signal |
switching signal; analogue signal; pulse signal; IO-Link; (configurable) |
Electrical design |
PNP |
Number of digital outputs |
2 |
Output function |
normally open / normally closed; (parameterisable) |
Max. voltage drop switching output DC [V] |
2 |
Permanent current rating of switching output DC [mA] |
250; (per output) |
Number of analogue outputs |
1 |
Analogue current output [mA] |
4…20; (scalable) |
Max. load [Ω] |
500 |
Pulse output |
consumed quantity meter |
Short-circuit protection |
yes |
Type of short-circuit protection |
pulsed |
Overload protection |
yes |
Measuring/setting range |
Measuring range |
Measuring/setting range; Flow monitoring; The values apply at: Ø 72 mm |
Measuring range |
0.5…143.9 m/s |
8…2110 m³/h |
0.12…35.18 m³/min |
|
Display range |
0…172.7 m/s |
0…2532 m³/h |
0…42.22 m³/min |
|
Resolution |
0.1 m/s |
2 m³/h |
0.02 m³/min |
|
Set point SP |
1.2…143.9 m/s |
18…2110 m³/h |
0.28…35.18 m³/min |
|
Reset point rP |
0.5…143.2 m/s |
6…2100 m³/h |
0.12…35 m³/min |
|
Analogue start point ASP |
0…107.9 m/s |
0…1582 m³/h |
0…26.38 m³/min |
|
Analogue end point AEP |
36…143.9 m/s |
528…2110 m³/h |
8.8…35.18 m³/min |
|
In steps of |
0.1 m/s |
2 m³/h |
0.02 m³/min |
|
Volumetric flow quantity monitoring |
Pulse value |
1…1000 x 10³ |
In steps of |
1 Nm³ |
Pulse length [s] |
0,7…2 (D = 72 mm) |
Temperature monitoring |
Measuring range [°C] |
0…60 |
Display range [°C] |
-12…72 |
Resolution [°C] |
0.2 |
Set point SP [°C] |
0.4…60 |
Reset point rP [°C] |
0.2…59.8 |
Analogue start point [°C] |
0…45 |
Analogue end point [°C] |
15…60 |
In steps of [°C] |
0.2 |
Accuracy / deviations |
Flow monitoring |
Repeatability [% of the measured value] |
± 1,5 |
Accuracy (in the measuring range) |
± (6 % MW + 0,6 % MEW); (D = 72mm, T = 22 °C; standard volume flow: 50…850 Nm³/h) |
Temperature monitoring |
Accuracy [K] |
± 2,5 (Q > 2 Nm³/h) |
Response times |
Flow monitoring |
Response time [s] |
0.1; (dAP = 0) |
Damping for the switching output dAP in steps [s] |
0 – 0,2 – 0,4 – 0,6 – 0,8 – 1 |
Temperature monitoring |
Dynamic response T05 / T09 [s] |
30 (Q > 2 Nm³/h) |
Software / programming |
Parameter setting options |
Flow monitoring; quantity meter; Preset counter; Temperature monitoring; hysteresis / window; normally open / normally closed; current/pulse output; display can be rotated and switched off; Display unit; totaliser |
Interfaces |
Communication interface |
IO-Link |
Transmission type |
COM2 (38,4 kBaud) |
IO-Link revision |
1.1 |
SDCI standard |
IEC 61131-9 |
Profiles |
Smart Sensor: Process Data Variable; Device Identification |
SIO mode |
yes |
Required master port type |
A |
Process data analogue |
3 |
Process data binary |
2 |
Min. process cycle time [ms] |
5 |
Supported DeviceIDs |
Type of operation |
DeviceID |
Default |
381 |
|
Operating conditions |
Ambient temperature [°C] |
0…60 |
Storage temperature [°C] |
-20…85 |
Max. relative air humidity [%] |
90 |
Protection |
IP 65 |
Mechanical data |
Weight [g] |
442 |
Materials |
PBT-GF20; PC; stainless steel (1.4301 / 304); FKM |
Materials (wetted parts) |
stainless steel (1.4401 / 316); stainless steel (1.4301 / 304); ceramics glass passivated; PEEK; polyester; FKM |
Process connection |
threaded connection G 1 internal thread |
Displays / operating elements |
Display |
Display unit |
4 x LED, green (Nm³/min, Nm³/h, Nm³, °C) |
function display |
1 x LED, green |
switching status |
2 x LED, yellow |
measured values |
alphanumeric display, 4-digit |
Sản phẩm cùng loại
Chat qua zalo
Để lại thông tin - Nhận giá ưu đãi
Gọi cho CÔNG NGHIỆP VÀ TỰ ĐỘNG HÓA COZIC để nhận tư vấn tốt nhất
Địa chỉ - CÔNG NGHIỆP VÀ TỰ ĐỘNG HÓA COZIC
78 Hoà Minh 1 - Hoà Minh - Liên Chiểu - Đà Nẵng - Việt Nam
Văn phòng - CÔNG NGHIỆP VÀ TỰ ĐỘNG HÓA COZIC
VP: 16 Vũ Huy Tấn - Hoà Hiệp Nam - Liên Chiểu- Đà Nẵng- Việt Nam
Thông báo sẽ tự động tắt sau 5 giây ...
| |