Product characteristics |
Number of inputs and outputs |
Number of analogue outputs: 1 |
Measuring range |
3.5…50 l/min |
0.29…4.145 m/s |
|
Process connection |
threaded connection G 3/4 DN15 |
Application |
Special feature |
Gold-plated contacts |
Measuring element |
1 x Pt 1000; (to DIN EN 60751, class B) |
Application |
for industrial applications |
Installation |
connection to pipe by means of an adapter |
Media |
water; glycol solutions; coolants |
Medium temperature [°C] |
-40…100 |
Min. bursting pressure [bar] |
25 |
Min. bursting pressure [MPa] |
2.5 |
Pressure rating [bar] |
12 |
Pressure rating [Mpa] |
1.2 |
Note on pressure rating |
|
Electrical data |
Operating voltage [V] |
8…33 DC |
Min. insulation resistance [MΩ] |
100; (500 V DC) |
Protection class |
III |
Power-on delay time [s] |
< 2 |
Inputs / outputs |
Number of inputs and outputs |
Number of analogue outputs: 1 |
Outputs |
Total number of outputs |
1 |
Output signal |
analogue signal |
Number of analogue outputs |
1 |
Analogue current output [mA] |
4…20; (water: Q [l/min] = 3,125 x (I – 4 mA); water-glycol: Q [l/min] = 3,125 x (I – 4 mA) – Qo see Figure 2) |
Max. load [Ω] |
< (Ub – 8 V) / 20 mA; Ub = 24 V: 800 |
Measuring/setting range |
Measuring range |
3.5…50 l/min |
0.29…4.145 m/s |
|
Temperature monitoring |
Internal heating temperature probe |
1 K/mW |
Measuring range [°C] |
-40…100 |
Accuracy / deviations |
Flow monitoring |
Accuracy (in the measuring range) |
Q < 50 % MEW: < 1 % MEW / Q > 50 % MEW: < 2 % MW; (water) |
Repeatability |
0,2; (% of the final value) |
Temperature monitoring |
Accuracy [K] |
± 0,3 ± 0,005 x T |
Response times |
Flow monitoring |
Response time [s] |
0.5 |
Operating conditions |
Ambient temperature [°C] |
-15…85 |
Note on ambient temperature |
medium temperature > 0 °C: -30…85 |
|
Storage temperature [°C] |
-30…85 |
Protection |
IP 65 |
Cavitation |
P(absolute) discharge / P(difference) > 5.5 to avoid cavitation |
Mechanical data |
Weight [g] |
76.5 |
Materials |
PA 6T |
Materials (wetted parts) |
ETFE; PA 6T; EPDM |
Tightening torque [Nm] |
12 |
Process connection |
threaded connection G 3/4 DN15 |
Remarks |
Remarks |
MW = measured value |
MEW = Final value of the measuring range |
|
Pack quantity |
1 pcs. |
Electrical connection |
Connection |
Connector: 1 x M12; Contacts: gold-plated |